DM-3700
●Sensors available for differential pressure, gauge pressure, pressure/vacuum, and absolute pressure.
Item | Description |
---|---|
Digital Display | 0000 ~ ±9999 |
Comparator Output | 5 outputs: Limits (HH、HI、LO、LL) and GO |
8 Channels | 8 individual programmable comparators |
Reading Holds | The reading can be held at any time. |
Peak value hold / Bottom value hold |
The peak value in pressure rise and the bottom value in pressure drop can be held. |
Auto-Zero | With the current value set to zero, pressure changes after that can be displayed. |
Analog Output | The analog voltage corresponding to the Pressure Range is output. (Amplifier optionally available) |
Digital Output | BCD output board is optionally available. Data can be transferred to external equipment. |
Keyboard Lock | The keyboard can be locked to prevent false operation. |
Standard | |||
Differential Pressure | Gauge Pressure | ||
Measuring Media | Air / Non-corrosive gas | Air / Non-corrosive gas / Liquid | |
Sensing Element | Beryllium copper | SUS630 | |
Transducer | Inductance type | Capacitance type | |
Accuracy*2 | Accuracy | ±0.15 % of F.S. ±1digit *1 | Temperature Characteristic at Zero Point |
±0.03 % of F.S./ °C | Span Temperature Characteristic |
±0.03 % of F.S./ °C |
Proof Pressure | Less than 50 kPa: 10 x F.S. | Less than 50 kPa: 10 x F.S. | 50 kPa or higher: 5 x F.S. | 10 MPa or higher: 1.5 x F.S. |
Maximum Line Pressure | 2 x Proof pressure | - | |
Response Time | 380 ms Max. (99 % of F.S.)*3 | 110 ms Max. (99 % of F.S.) | |
Digital Display | 0000 ~ ±9999 | ||
Sampling Time | 200 ms | ||
Comparator Output | HH, HI, GO, LO, LL (Relay contact output) | ||
Analog Output | The analog voltage 1, 2 or 5 VDC corresponding to the Pressure Range is output. (5 and 10 VDC are optionally available.) Option (Open collector, TTL) |
||
BCD Output | Option (Open collector, TTL) | ||
Power Source | 100 to 240 VAC multi-power supply ±10 %, 50/60 Hz 1.0 A | ||
Operating Temperature | 5 to 40 °C | ||
Pressure Inlet Port | Rc1/8 | ||
Size | 184(W)×108 (H)×232 (D) mm | ||
Weight | 3.5 kg | ||
Panel-Cut Size | 181.5(W)×101(H) mm |
*1. Accuracy of drip-proof model: ±0.25% of F.S.±1digit
*2. Linearity, repeatability, and hysteresis are included.
*3. The response time in the range of 0 to 200 Pa is approx. 540 ms.
Standard
Range Code |
Sensor | Measurement Range |
Range Code |
Sensor | Measurement Range |
2PD | Differential pressure | 0 to 200 Pa | 5MG | Gauge pressure | 0 to 5 MPa |
5PD | Differential pressure | 0 to 500 Pa | 10MG | Gauge pressure | 0 to 10 MPa |
1KD | Differential pressure | 0 to 1 kPa | 20MG | Gauge pressure | 0 to 20 MPa |
2KD | Differential pressure | 0 to 2 kPa | 50MG | Gauge pressure | 0 to 50 MPa |
5KD | Differential pressure | 0 to 5 kPa | V2PD | Vacuum differential pressure | -200 Pa to 0 |
10KD | Differential pressure | 0 to 10 kPa | V5PD | Vacuum differential pressure | -500 Pa~0 |
20KD | Differential pressure | 0 to 20 kPa | V1KD | Vacuum differential pressure | -1 kPa to 0 |
50KD | Differential pressure | 0 to 50 kPa | V2KD | Vacuum differential pressure | -2 kPa to 0 |
50KG | Gauge pressure | 0 to 50 kPa | V5KD | Vacuum differential pressure | -5 kPa to 0 |
100KD | Differential pressure | 0 to 100 kPa | V10KD | Vacuum differential pressure | -10 kPa to 0 |
100KG | Gauge pressure | 0 to 100 kPa | V20KD | Vacuum differential pressure | -20 kPa to 0 |
200KG | Gauge pressure | 0 to 200 kPa | V50KD | Vacuum differential pressure | -50 kPa to 0 |
500KG | Gauge pressure | 0 to 500 kPa | V50KG | Vacuum gauge pressure | -50 kPa to 0 |
1MG | Gauge pressure | 0 to 1 MPa | V100KD | Vacuum differential pressure | -100 kPa to 0 |
2MG | Gauge pressure | 0 to 2 MPa | V100KG | Vacuum gauge pressure | -100 kPa to 0 |
Pressure/Vacuum
Range Code |
Sensor | Measurement Range |
Range Code |
Sensor | Measurement Range |
X2PD | Differential pressure | 0 to ±200 Pa | X50KD | Differential pressure | 0 to ±50 kPa |
X5PD | Differential pressure | 0 to ±500 Pa | X50KG | Gauge pressure | 0 to ±50 kPa |
X1KD | Differential pressure | 0 to ±1 kPa | X100KD | Differential pressure | 0 to ±100 kPa |
X2KD | Differential pressure | 0 to ±2 kPa | X100KG | Gauge pressure | 0 to ±100 kPa |
X5KD | Differential pressure | 0 to ±5 kPa | X200KG | Gauge pressure | -100 to 200 kPa |
X10KD | Differential pressure | 0 to ±10 kPa | X500KG | Gauge pressure | -100 to 500 kPa |
X20KD | Differential pressure | 0~±20 kPa |
Since the lowest calibration point for all the Cosmo products is -90 kPa, the guaranteed range is down to -90
Model:DM-3700(A.BC.D)
A, B and C in the model notation are mandatory.
Category | Code | Description | |
---|---|---|---|
A | Pressure Range | See the Range List | |
B | Sensor | N | Internal pressure sensor (Standard) |
S1 | External pressure sensor | ||
S2 | External pressure sensor with drip-proof (Gauge pressure only) *1 | ||
C | Sensor Cable Length | L1 | 3m (Standard) |
LX1 | 1m | ||
LX5 | 5m | ||
LX10 | 10m | ||
BCD | D1 | Open collector | |
D2 | TTL output | ||
Analog Output | F1 | Analog output F.S. 5 V | |
F2 | Analog output F.S. 10 V | ||
Display Digits | Q | Change of the minimum display digit (Only for the ranges that include 1)*2 | |
Mounting Brackets | P | Mounting Brackets | |
D | Power Cord (Standard accessory) |
VA | 125 VAC 3m power cord |
VE | 250 VAC 2 m power cord | ||
VK | 250 VAC 2 m power cord (Mandatory for Chinese customers) |
*1. Accuracy of S2 model: ±0.25% of F.S.±1digit.
*2. The analog output is amplified to 10 V.